Purpose: For interventional radiology, dose management has persisted as a crucially important issue to reduce radiation risks to patients and medical staff. This study examined the concept and feasibility of a real-time dose visualization system using mixed reality technology for dose management in interventional radiology.
Methods: An earlier reported Monte-Carlo-based real-time dose estimation system for interventional radiology was adopted for our system. We also developed a system of acquiring fluoroscopic conditions to input them into the Monte Carlo system. Then, we combined the Monte Carlo system with the wearable device for three-dimensional holographic visualization. The estimated doses were transferred sequentially to the device. The patient's dose distribution was then projected on the patient body. The visualization system also has a mechanism to detect one's position in a room to estimate the user's exposure dose based on the positional information. The dose was visualized continuously in front of the user.
Results: An end-to-end system test was performed using a human phantom. The acquisition system accurately recognized conditions that were necessary for real-time dose estimation. The dose hologram represented the patient dose. The user dose was changed correctly depending on conditions and positions.
Conclusion: Mixed reality dose visualization is expected to improve exposure dose management for patients and physicians by exhibiting invisible radiation exposure in real space.
Funding Support, Disclosures, and Conflict of Interest: This work was partly supported by a Japan Society for the Promotion of Science (JSPS) KAKENHI Grant (No. 18K07646).