Click here to

Session: [Return to Session]

Characterization of a Novel Kilovoltage Fan Beam CT System Integrated with a Ring Gantry Linac

G Gibbard1*, J Tan1, T Zhuang1, Y Park1, K Wang1, W Lu1, M Lin1, R Reynolds1, A Godley1, A Pompos1, S Jiang1, B Cai1, (1) UT Southwestern Medical Center, Dallas, TX,

Presentations

(Saturday, 3/26/2022)   [Central Time (GMT-5)]

Purpose: This study aims to characterize the performance of the kV Fan-Beam Computed Tomography (kV-FBCT) imaging system on RefleXion X1 by use of national standards (TG40, TG142, TG148, TG179) and machine specific tolerances. RefleXion X1 is a novel radiotherapy machine that utilizes the kV-FBCT imaging system for IMRT, SBRT and Biology-guided Radiotherapy (BgRT). The kV-FBCT is used to guide patient setup and target localization.

Methods: Multiple imaging protocols were supported on this platform, tests were performed with a standard imaging protocol of 120 kV/1 s/150 mA/ 0.5 pitch/1.25 mm and 16 slice helical acquisitions. Standard image quality metrics, including slice thickness accuracy, Modulation Transfer Function (MTF) at 50%, spatial resolution, low-contrast resolution, geometric accuracy, contrast scale, CT Hounsfield unit (HU) number accuracy, regional uniformity & noise, and kV-MV Isocenter were evaluated using an image quality phantom (Catphan 500, The Phantom Laboratory, New York, NY), a water phantom (Jaszczak phantom) and Ball-Cube-II and Head Phantom.

Results: All major image quality metrics were within either the national guideline or vendor-recommended tolerances. The slice thickness, MTF at 50%, spatial resolution were measured to be 1.69 mm, 4.51 lp/cm, 9 lp/cm, respectively. The clinical spatial resolution and low contrast visibility was found to be acceptable with greater than 4 lp/cm and the smallest visible target less than 5 mm. The geometric accuracy tests passed within 1 mm at a 50 mm distance. The CT number uniformity is within 2.7 HU, contrast scale difference is within 0 HU. The CT number accuracy and noise are all within 15 HU. The total targeting error was less than 1 mm.

Conclusion: Comprehensive characterization of the kV-FBCT has been shown to provide acceptable image quality for image guidance. The image unit onboard the RefleXion X1 meets specifications and satisfies requirements for routine clinical use.

ePosters

Keywords

PET, Helical CT, Image-guided Therapy

Taxonomy

IM- Multi-Modality Imaging Systems: CT/PET - human

Contact Email