Purpose: External beam radiotherapy can deliver a highly conformal dose to targets, with steep dose falloff to organs-at risk. Nonetheless, the actual delivered dose distribution will differ from planned dose without in-vivo dose verification during treatment. Electronic portal imaging devices (EPIDs) have been used as a dosimetric tool for pre-treatment and in-vivo patient-specific verification. In this study, a deep learning-based method has been developed to reconstruct in-vivo dose using EPID signals.
Methods: A deep learning-based reconstruction algorithm was utilized to learn a mapping from an EPID dose signal simulated by Monte Carlo method to in-vivo dose distribution at different distance from EPID. A ResNet architecture was trained to reproduce patient dose distribution from the input consisting of a CT reconstruction and a simulated EPID dose signal. Finally, the in-vivo dose distribution was compared with the planned dose to evaluate the quality of prediction.
Results: Mean absolute percentage error (MAPE) and a gamma analysis were quantified between the in-vivo dose distribution and the planned dose. Both comparisons indicate that the reconstructed in-vivo dose was in a good agreement with the original dose.
Conclusion: EPID-based 3D in vivo dosimetry reconstruction provides an efficient approach to monitor the accuracy of dose delivery during treatment.