Purpose: Experimental measurements and validation of 2D dose rate distributions in proton pencil beam scanning (PBS) FLASH radiation therapy (RT) are currently lacking. In this work, we characterize a newly designed 2D strip-segmented ionization chamber array (SICA) with high spatial and temporal resolution and demonstrate its applications in a modern proton PBS delivery system for FLASH-RT.
Methods: A dedicated research beamline of the Varian ProBeam system was employed to deliver a 250 MeV proton PBS beam with nozzle currents up to 215 nA. The spatial, temporal, and dosimetric performance of the SICA was characterized and compared with measurements using a parallel-plate ion chamber (PTW Advanced Markus chamber), Gafchromic films (EBT-XD), and a Faraday Cup. A novel reconstruction approach for 2D strip ionization chamber was proposed and enabled the first experimental measurements of the 2D dose rate distributions under a proton PBS FLASH-RT delivery using the SICA.
Results: The SICA demonstrated a position accuracy of 0.12 ± 0.02 mm at a 20 kHz sampling rate (50 μs per event) and a linearity of R² > 0.99 for both dose and dose rate with respect to nozzle beam currents ranging from about 1 nA to 215 nA. The 2D dose comparison to the film measurement resulted in a gamma passing rate of 99.8% (2 mm/2%). The measured proton PBS 2D FLASH dose rate distribution was compared to simulation results and showed a gamma passing rate of 97.3% (2 mm/2%).
Conclusion: The newly designed SICA demonstrated excellent spatial, temporal, and dosimetric performance and is well suited for commissioning, quality assurance (QA), and a wide range of clinical applications in proton PBS FLASH radiotherapy.
Protons, Quality Assurance, Ionization Chamber
TH- External Beam- Particle/high LET therapy: Proton therapy – quality assurance