ePoster Forums
Purpose: To evaluate the influence of plan modulation on IMRT/VMAT dose calculation sensitivity to changes in MLC parameter setting.
Methods: A total of 49 plans(24 IMRT,25 VMAT) were selected for dosimetric evaluation, with each plan utilizing TrueBeam delivery/6x mode. The cases covered a wide range of disease sites-including pelvis, thorax, brain, and HN. The Modulation Factor(MF) was defined as Plan MU divided by Rx Dose. For IMRT cases-the MF range was [169, 1439]; for VMAT cases-the MF range was [146, 628]. To eliminate effects from tissue heterogeneity, each plan was calculated on water-equivalent phantom dataset using both AAA and AcurosXB algorithms. For each algorithm, multiple DLG settings were utilized to evaluate the impact of MF on dose calculation/DLG sensitivity. For each plan, the point dose(at isocenter or other meaningful high-dose location)was used for dose sensitivity evaluation.
Results: Each plan was calculated using 3 DLG settings (0.110cm,0.135cm,and 0.160cm), with dose sensitivity defined as percent difference in dose between the highest and lowest DLG setting. The plans were grouped into 3 tiers based on MF value: Low(100
Conclusion: Our study offers some insight into the behavior of IMRT/VMAT dose calculation across a spectrum of plan modulation values. Given that DLG sensitivity increases with modulation and delivery modality choice(IMRT vs. VMAT), our results suggest that plan selection during algorithm validation/commissioning should cover a useful range of modulation factors for both delivery modalities. In addition, we note that these trends hold for both algorithms(AAA and AcurosXB) tested in this study.
Not Applicable / None Entered.
Not Applicable / None Entered.